Manufacturer Product Updates
SUN NUCLEAR IVD SOLUTIONS SOFTWARE VERSION 3.02.00 UPDATE - 10/2/10
We are pleased to inform you that version 3.02.00 of the Sun Nuclear IVD software is now available for your download. Please visit the Sun Nuclear website and download the software at your convenience.
http://www.sunnuclear.com/login/login.asp?LinkTo=/support/download/downloads.asp
The software upgrade will bring significant improvements in flexibility, time saving and ease of use.
If you need any help with the download please do not hesitate to contact us.
The new IVD Solutions software version 3.02.00 has been released and it is available on the IVD Solutions CD and the website. Below are all the changes implemented in this release.
Version 3.02.00
New Features:
1) Create and modify your own correction factors in addition to using the default factors provided.
2) Print a single measurement or a patient measurement to a PDF file.
3) Import patient data exported by RadCalc in a RTF file format.
4) Logged in user is displayed at the bottom of the screen.
5) Database version is displayed at the bottom of the screen.
6) Correction factors set can be copied to other diodes.
7) Two alarm levels, one to alert physics of deviation, one to alert therapists of harm.
8) Expected dose information can be added after measurement.
9) Improved main screen and tab layout.
10) New 'thickness' default correction factor.
11) Interface to ARIA through RVP software to export an ARIA readable XML format file.
12) Export an ARIA readable XML format file on demand without using the ARIA/RVP interface.
13) Electronic sign off of patient measurement results by fraction.
14) Average of two diode measurements can now be added to the patient printout.
15) Added patient ID to the patient name on the patient dose collection screen.
16) Measured by and measured time/date added to the diode display in the Patient Plans tab on the field screen.
Fixed Defect:
1) Changes to a pod number were not registering in the database.
2) When clicking on an incomplete field, an access violation was occurring.
3) A completed patient with deleted fields was still showing as 'incomplete' after all fields were measured.
4) The scrollbar was causing an 'index out of range' error when using greater than 4 pods.
5) When adding a patient, if the ID was entered before the patient name, the “patient ID number already used” message appeared spuriously.
6) When entering a diode serial number, the software would allow entry of all the digits.
7) Temperature compensation values could be entered for positive diodes which cannot be compensated.
8) Clicking on the error column and then clicking on the setup button caused the interface to crash.
9) Background corrected counts were not being displayed in the calibration screen.
|